Test ID BLOD0670 Dilute Russell Viper Venom Time

AKA

DRVVT

Specimen Type/Requirements

Blue top (Sodium Citrate) tube - Plasma, Double Spun Platelet Poor

 

See Coagulation Testing and Platelet Poor Plasma Preparation

 

The 3.2% Sodium Citrate tube must be filled to appropriate volume to acquire the 9:1 blood to anticoagulant ratio that is critical for valid results.

 

Do not store in a frost-free freezer.  Do not test specimens that have gone through freeze/thaw cycles.

 

Test is not affected by hemolysis or lipemia.

Specimen Volume

 Preferred Volume     2.0 mL   
 Minimum Volume     1.0 mL   

 

Stability/Transport

 Room Temperature     4 hours     
 Refrigerated     Not Acceptable     
 Frozen     14 days    Preferred for transport   

 

Performed Test Frequency

Tuesday and Friday

Additional Information

Abnormal results can be obtained in patients on anticoagulant therapy, including the newer agents (e.g. rivaroxaban, apixaban, etc.). This can result in false interpretation of lupus inhibitor.

Methodology

Stagos Instrument

Performing Lab

Sanford Laboratories Sioux Falls

CPT

85613