Sign in →

Test ID BLOD0555 Chloride

AKA

CL

Specimen Type/Requirements

Gold top (Serum w/gel) tube - Serum (preferred)

Red top (Serum w/out gel) tube - Serum (preferred)
Light Green top (Lithium Heparin w/gel) tube - Plasma

 

Serum/plasma should be separated from cells within 4 hours of collection.

 

Test is not affected by hemolysis or lipemia.

Specimen Volume

 Preferred Volume     1.0 mL   
 Minimum Volume     0.6 mL   

 

Stability/Transport

 Room Temperature     7 days     
 Refrigerated     7 days     Preferred for transport   
 Frozen     1 year     

 

Performed Test Frequency

Monday through Sunday

Methodology

Sioux Falls – ISE Indirect
Rapid City – ISE Indirect

Performing Lab

Sanford Laboratories Sioux Falls
Sanford Laboratories Rapid City

CPT

82435