Test ID BLOD0150 Aluminum, Serum
Useful For
Monitoring for aluminum toxicity in patients undergoing dialysis
Routine aluminum screening
Monitoring metallic prosthetic implant wear
Specimen Type
SerumSpecimen Required
Patient Preparation: High concentrations of gadolinium and iodine are known to potentially interfere with most inductively coupled plasma mass spectrometry-based metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Supplies:
-Metal Free B-D Tube (No Additive), 6 mL (T184)
-Metal Free Specimen Vial (T173)
Container/Tube: 6-mL Plain, royal blue-top Vacutainer plastic trace element blood collection tube
Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial
Specimen Volume: 1.2 mL
Collection Instructions: See Metals Analysis Specimen Collection and Transport for complete instructions.
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | METAL FREE |
Ambient | 28 days | METAL FREE | |
Frozen | 28 days | METAL FREE |
Day(s) Performed
Monday through Friday
Report Available
1 to 4 daysMethod Name
Triple-Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS/MS)
Performing Laboratory

CPT Code Information
82108
Reporting Name
Aluminum, SReference Values
<7 ng/mL
<60 ng/mL (dialysis patients)
For International System of Units (SI) conversion for Reference Values, see www.mayocliniclabs.com/order-tests/si-unit-conversion.html
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.Forms
If not ordering electronically, complete, print, and send a Renal Diagnostics Test Request (T830) with the specimen.
Special Instructions
Secondary ID
8373SANFORD LABORATORY INTERFACE BUILD INFORMATION
Result Code | Result Code Description |
---|---|
302 | Aluminum |